==Post-operative evaluation==
 
==Post-operative evaluation==
File:X-ray of knee prosthesis with overhang, annotated.jpg|thumb|150px|Overhang (arrow) does not seem to have any detrimental effect.<ref>{{cite journal|title=The Impact Of Tibial Component Overhang On Outcome Scores And Pain In Total Knee Replacement|date=2018-02-21|author=S.G.F. Abram, A.G Marsh, F. Nicol, A.S. Brydone, A. Mohammed, S.J. Spencer|journal=Orthopaedic Proceedings}}</ref> It is therefore not necessary to measure or report.]]
+
[[File:X-ray of knee prosthesis with overhang, annotated.jpg|thumb|150px|Overhang (arrow) does not seem to have any detrimental effect.<ref>{{cite journal|title=The Impact Of Tibial Component Overhang On Outcome Scores And Pain In Total Knee Replacement|date=2018-02-21|author=S.G.F. Abram, A.G Marsh, F. Nicol, A.S. Brydone, A. Mohammed, S.J. Spencer|journal=Orthopaedic Proceedings}}</ref> It is therefore not necessary to measure or report.]]
 
Knee replacement is evaluated by the following measures:
 
Knee replacement is evaluated by the following measures:
 
<gallery mode=packed heights=380>
 
<gallery mode=packed heights=380>
 
File:X-ray of HKA angle with knee prosthesis.jpg|'''HKA''': Hip-knee-ankle angle, which is ideally between 3° [[genu varum|varum]] to 3° [[Genu valgum|valgum]] from a [[right angle]].<ref name=Inui2013>{{cite journal|last1=Inui|first1=Hiroshi|last2=Taketomi|first2=Shuji|last3=Nakamura|first3=Kensuke|last4=Takei|first4=Seira|last5=Takeda|first5=Hideki|last6=Tanaka|first6=Sakae|last7=Nakagawa|first7=Takumi|title=Influence of navigation system updates on total knee arthroplasty|journal=Sports Medicine, Arthroscopy, Rehabilitation, Therapy & Technology|volume=5|issue=1|pages=10|year=2013|issn=1758-2555|doi=10.1186/2052-1847-5-10|pmid=23638774|pmc=3653716}} [https://creativecommons.org/licenses/by/2.0 (CC-BY-2.0)]</ref>
 
File:X-ray of HKA angle with knee prosthesis.jpg|'''HKA''': Hip-knee-ankle angle, which is ideally between 3° [[genu varum|varum]] to 3° [[Genu valgum|valgum]] from a [[right angle]].<ref name=Inui2013>{{cite journal|last1=Inui|first1=Hiroshi|last2=Taketomi|first2=Shuji|last3=Nakamura|first3=Kensuke|last4=Takei|first4=Seira|last5=Takeda|first5=Hideki|last6=Tanaka|first6=Sakae|last7=Nakagawa|first7=Takumi|title=Influence of navigation system updates on total knee arthroplasty|journal=Sports Medicine, Arthroscopy, Rehabilitation, Therapy & Technology|volume=5|issue=1|pages=10|year=2013|issn=1758-2555|doi=10.1186/2052-1847-5-10|pmid=23638774|pmc=3653716}} [https://creativecommons.org/licenses/by/2.0 (CC-BY-2.0)]</ref>
 
File:Postoperative X-ray of normal knee prosthesis, anteroposterior view, annotated.jpg|- '''FFC''': frontal femoral component angle. It is typically regarded as optimal when being 2–7° in valgus.<ref name="GromovKorchi2014">{{cite journal|last1=Gromov|first1=Kirill|last2=Korchi|first2=Mounim|last3=Thomsen|first3=Morten G|last4=Husted|first4=Henrik|last5=Troelsen|first5=Anders|title=What is the optimal alignment of the tibial and femoral components in knee arthroplasty?|journal=Acta Orthopaedica|volume=85|issue=5|year=2014|pages=480–487|issn=1745-3674|doi=10.3109/17453674.2014.940573|pmid=25036719|pmc=4164865}}</ref><br>- '''FTC''': frontal tibial component angle, which is regarded as optimal when being at a [[right angle]]. A varus position of more than 3° has generally been found to increase the failure rate of the prosthesis.<ref name="GromovKorchi2014"/>
 
File:Postoperative X-ray of normal knee prosthesis, anteroposterior view, annotated.jpg|- '''FFC''': frontal femoral component angle. It is typically regarded as optimal when being 2–7° in valgus.<ref name="GromovKorchi2014">{{cite journal|last1=Gromov|first1=Kirill|last2=Korchi|first2=Mounim|last3=Thomsen|first3=Morten G|last4=Husted|first4=Henrik|last5=Troelsen|first5=Anders|title=What is the optimal alignment of the tibial and femoral components in knee arthroplasty?|journal=Acta Orthopaedica|volume=85|issue=5|year=2014|pages=480–487|issn=1745-3674|doi=10.3109/17453674.2014.940573|pmid=25036719|pmc=4164865}}</ref><br>- '''FTC''': frontal tibial component angle, which is regarded as optimal when being at a [[right angle]]. A varus position of more than 3° has generally been found to increase the failure rate of the prosthesis.<ref name="GromovKorchi2014"/>
File:Postoperative X-ray of normal knee prosthesis, lateral view, annotated.jpg|- '''Anterior femoral notching''' (the femoral component causing reduced thickness of the distal femur anteriorly), seems to cause an increased risk of fractures when exceeding about 3 mm.<ref name="LeeWang2015">{{cite journal|last1=Lee|first1=Ju Hong|last2=Wang|first2=Seong-Il|title=Risk of Anterior Femoral Notching in Navigated Total Knee Arthroplasty|journal=Clinics in Orthopedic Surgery|volume=7|issue=2|year=2015|pages=217|issn=2005-291X|doi=10.4055/cios.2015.7.2.217}}</ref><br>- '''LTC''': lateral (or sagittal) tibial component angle, which is ideally positioned so that the tibia is 0–7° flexed compared to at a right angle with the tibial plate.<ref name="GromovKorchi2014"/>]]
+
File:Postoperative X-ray of normal knee prosthesis, lateral view, annotated.jpg|- '''Anterior femoral notching''' (the femoral component causing reduced thickness of the distal femur anteriorly), seems to cause an increased risk of fractures when exceeding about 3 mm.<ref name="LeeWang2015">{{cite journal|last1=Lee|first1=Ju Hong|last2=Wang|first2=Seong-Il|title=Risk of Anterior Femoral Notching in Navigated Total Knee Arthroplasty|journal=Clinics in Orthopedic Surgery|volume=7|issue=2|year=2015|pages=217|issn=2005-291X|doi=10.4055/cios.2015.7.2.217}}</ref><br>- '''LTC''': lateral (or sagittal) tibial component angle, which is ideally positioned so that the tibia is 0–7° flexed compared to at a right angle with the tibial plate.<ref name="GromovKorchi2014"/>
 
</gallery>
 
</gallery>
  

Revision as of 13:31, 2 July 2019

Author: Mikael Häggström [notes 1]

Post-operative evaluation

File:X-ray of knee prosthesis with overhang, annotated.jpg
Overhang (arrow) does not seem to have any detrimental effect.[1] It is therefore not necessary to measure or report.

Knee replacement is evaluated by the following measures:

Report

In Swedish healthcare[5], the numbers of the angles are not reported if being within normal limits. Example of an normal case:

  • Postoperative images of knee implant in unremarkable position.

Also, unless you know the referring physician is familiar with the same angle terminology, preferably report deviating angles as varus/valgus deviation or ventral/dorsal angulation of either:

  • The femur component in relation to femoral diaphysis
  • The tibial diaphysis in relation to the tibial component.
See also: General notes on reporting

Follow-up

Potential complications that need to be evaluated on follow-up are as follows.

Radiolucent lines

Radiolucent lines may indicate loosening of the implant. A radiolucent line thinner than than 2 mm can be tolerated at the cement-bone interface (for cemented implants) or implant-bone interface (for cementless implants) if it remains stable and appears within the first 6 months (cemented implants) or the first 2 years (cementless implants) after surgery.[6]

Displacement

A change in position is indicated by a significant change in either of the angles listed in the post-operative evaluation section above.[6]

Infection

Signs of infection include:[6]

  • Rapidly progressing radiolucency and/or osteolysis
  • Periosteal reaction
  • Bubbles of air within soft tissue or fluid collection.

Other complications

  • Fractures of prosthesis components.
  • Wear of the polyethylene insert, which is the case when progressive thinning occurs over time.
  • Dissociation of the polyethylene insert from the patellar component or tibial baseplate.

Reporting

In Swedish healthcare[7], an example report of a normal case may be:

  • No changes since previous images on <date>.
See also: General notes on reporting

Notes

  1. For a full list of contributors, see article history. Creators of images are attributed at the image description pages, seen by clicking on the images. See Radlines:Authorship for details.

References

  1. S.G.F. Abram, A.G Marsh, F. Nicol, A.S. Brydone, A. Mohammed, S.J. Spencer (2018-02-21). "The Impact Of Tibial Component Overhang On Outcome Scores And Pain In Total Knee Replacement ". Orthopaedic Proceedings. 
  2. Inui, Hiroshi; Taketomi, Shuji; Nakamura, Kensuke; Takei, Seira; Takeda, Hideki; Tanaka, Sakae; Nakagawa, Takumi (2013). "Influence of navigation system updates on total knee arthroplasty ". Sports Medicine, Arthroscopy, Rehabilitation, Therapy & Technology 5 (1): 10. doi:10.1186/2052-1847-5-10. ISSN 1758-2555. PMID 23638774.  (CC-BY-2.0)
  3. 3.0 3.1 3.2 Gromov, Kirill; Korchi, Mounim; Thomsen, Morten G; Husted, Henrik; Troelsen, Anders (2014). "What is the optimal alignment of the tibial and femoral components in knee arthroplasty? ". Acta Orthopaedica 85 (5): 480–487. doi:10.3109/17453674.2014.940573. ISSN 1745-3674. PMID 25036719. 
  4. Lee, Ju Hong; Wang, Seong-Il (2015). "Risk of Anterior Femoral Notching in Navigated Total Knee Arthroplasty ". Clinics in Orthopedic Surgery 7 (2): 217. doi:10.4055/cios.2015.7.2.217. ISSN 2005-291X. 
  5. NU Hospital Group, Sweden, Sep 2018
  6. 6.0 6.1 6.2 Cyteval, C. (2016). "Imaging of knee implants and related complications ". Diagnostic and Interventional Imaging 97 (7-8): 809–821. doi:10.1016/j.diii.2016.02.015. ISSN 22115684. 
  7. NU Hospital Group, Sweden, Oct 2018