Difference between revisions of "Fluoroscopy of central venous catheters"

From radlines.org
Jump to navigation Jump to search
(Started)
(No difference)

Revision as of 09:13, 4 July 2018

Author: Mikael Häggström [notes 1]

The fibrin sheath is indirectly seen in these images before and after radiocontrast infusion, as the radiocontrast collects around the catheter.

Fibrin sheaths of central venous catheters can be suspected when it is possible to infuse into it but not to aspirate.

Technique

Fibrin sheaths are usually detected by fluoroscopic catheter contrast angiography.[1] Non-ionic contrastwith a high iodine concentration (for example 300 mg/ml) is injected by hand using a 20-ml syringe. A precontrast view is followed by a series of images with at 2–4 images/s and small field-of-view.[1]

References

  1. 1.0 1.1 Knutstad, K.; Hager, B.; Hauser, M. (2016). "Radiologic diagnosis and management of complications related to central venous access ". Acta Radiologica 44 (5): 508–516. doi:10.1080/j.1600-0455.2003.00115.x. ISSN 0284-1851. 


Cite error: <ref> tags exist for a group named "notes", but no corresponding <references group="notes"/> tag was found, or a closing </ref> is missing